RIGOL
© Copyright
RIGOL
Technologies, Inc. 2007
2-75
User’s Guide for DG1000 Series
1
.
View the last message display
View the last message display.
2
.
View the remote Command error queue
View the remote command error queue in the remote operation.
3
.
Get HELP on any key
To get context-sensitive help on any front-panel key or menu softkey, press and hold
that key for a few times.
4
.
Generate a DC-only Voltage level
(1)
Press Utility and select DC On.
(2)
Press Offset, then enter the desired voltage level.
(3)
Choose DC On/Off
Æ
DC Off, or press any function key to resume the normal
waveform operation.
5
.
Generate a modulated Waveform
(1)
Configure the Carrier waveform by selecting the waveform shapes, frequency,
amplitude, etc.
(2)
Configure the modulating waveform by pressing the Mod key.
(3)
To change the Carrier parameter, press the lighted function key (Sine , Square
etc).
(4)
To turn off the modulation, press the lighted Mod key.
6
.
Create an arbitrary waveform
(1)
Press Arb and choose Edit
Æ
Create.
(2)
Enter the desired period, voltage limit, and the number of points and then press
“Edit Pt”, the first two points have defined for you.
(3)
Rotate the knob to select the desired point for editing and then use Time and
Voltage to define the point. Press Insert to insert additional points as needed.
(4)
The waveform editor connects the last point to the voltage level of point #1 to
create a continuous waveform. Make sure that the time value of the last point
less than the specified period.
(5)
Press Save to store the waveform and exit the waveform editor.