RIGOL
© Copyright
RIGOL
Technologies, Inc. 2007
User’s Guide for DG1000 Series
1-13
To Set Trigger/Output
As shown in Figure 1-21, there are two buttons on the right side of the operation
panel, which are used to set Trigger and Output Control. The instruction below will
help you to be familiar with these functions.
Figure 1-21
Trigger/ Output Button
1.
Press Trig/Run Button, choose internal/ external or manual Trigger (Manual
Trigger can only be used in Sweep and N-Cycle Burst)
The default setting for Trigger is “Internal”. In this mode, when the Sweep or
Burst Mode is also selected, the Generator will continuously generate burst. At
this time, press Trig/Run button, the instrument will shift from the “Automatic”
Trigger mode into “Manual” Trigger mode.
When the generator uses the” External” Trigger Mode, if the Sweep or the Burst
Mode is selected, signal will be continuously generated. At this time, press
Trig/Run button, the instrument state will not change, and it will show “The
instrument has already been triggered”.
Every time you press the Trig/Run button, “Manual” Trigger will start a sweep or
generate a burst. Press the button again, and the generator will be triggered
again.
2.
Press Output Button, activate or deactivate the output signal.
If an overload message is shown, disconnect the external equipment from the
output terminals and press Output button, reactivate the output terminal.