Example 3: To Generate a Ramp Wave
How to generate a Ramp waveform with Period of 10ms, Amplitude of 100mV
PP
,
Offset of 20mV
DC
and Symmetry of 80%.
The operation steps:
1.
Set the Period.
(1)
Press Ramp
Æ
Freq/Period and choose Freq which will display in reverse color.
(2)
Input “10” from the keypad and choose the unit “s” to set the Period as 10ms.
2.
Set the Amplitude.
(1)
Press Ampl/ HiLev and choose Ampl which will display in reverse color.
(2)
Enter “100” from the keypad and choose the unit “mV
PP
” to set the Amplitude as
100mV
PP
.
3.
Set the Offset.
(1)
Press Offset /LoLev to choose Offset which will display in reverse color.
(2)
Enter “20” from the keypad and choose the unit “mV
DC
” to set the Offset as
20mV
DC
.
4.
Set the Symmetry.
(1)
Press Symm which will display in reverse color.
(2)
Enter “80” from the keypad and choose the unit “%” to set the Symmetry as
80%.
When the Frequency, Amplitude Offset and Symmetry were set, the generated
waveform is shown in Figure 3-3.
Figure 3-3 Ramp Waveform Output