User’s Guide for DG1000 Series
3-8
6.
Edit the Waveform Points.
Edit the Voltage and Time of the points to define a waveform. Insert or delete
the point If needed.
Press Points, use the knob or the keypad to shift the point number. The voltage and
time of the points are defined below in Table 3-1.
Table 3-1 The Setting of Time and Voltage of the points
Point Time Voltage
1 0s 0V
2 4
μ
s 4V
3 8
μ
s 0V
4 10
μ
s -2V
7.
To save the Waveform.
Press Save, to store the waveform in one of the 4 Non-Volatile Memories.
Press Store, to store the finished waveform in the Volatile Memory. Only one
self-built waveform can be stored. If a new one is stored, the old one will be
erased.
The waveform is generated through the steps above. The waveform is shown in
Figure 3-8.
Figure 3-8 The Self-Defined Ramp Waveform