User’s Guide for DG1000 Series
2-30
To set the Trig Level
The trig level (-3V
~
+3V) is divided into share, each “0.1” is 6 mV, it means the
adjustment interval is 6 mV. For example, the input is “62.0”, then the trig level is set
to be : -3V + (62.0 / 0.1) × 6 mV
=
0.72V.
(1)
Press Count
Æ
Setup
Æ
TrigLev , to set the trig level.
The trig level display on the screen is the default value when power on or the value
set at the last time. If the current value is effective, it could be use immediately.
(2)
Input the required trig level.
Use the digits keyboard of the knob, input the required parameter, the counter will
immediately adjust the trig level, and trigger according to the specified value.
Figure 2-42 To set the trig level
Note: In the DC coupling mode, users are need to adjust the trig level
manually.