4-154 Menus and Commands
6-00058-05 Rev A
Library Submenu
Path:
Main Menu
Service
Diags
Library
Use the Library Submenu to perform move tests. See Figure 4-
239.
Figure 4-239
Library Submenu
Storage Move
Path:
Main Menu
Service
Diags
Library
Storage
Move
Use Storage Moves to move a cartridge through every storage
cell in a given range. The cartridge is moved to or from the I/
E station and then to the specified cells. This test requires a
service-level password. After a password is provided, the
system prompts for the I/E station where the source cartridge
is located. Then, the system prompts for the beginning and
ending storage location. The test can be cancelled at any. See
Figure 4-240.
Figure 4-240
Storage Moves Dialog
Parameters
Description
Cell
Beginning and ending storage
locations
Status
Shows the status of the test
(Running, Canceled)
Cancel
Y
to cancel the test
N
to continue running the test
>
Storage Move
...
Corner Moves...
STORAGE MOVE TEST
Cell: 01 A 2 10
Status: Running
Cancel: N
Note
If the operation
stops, the
Cancel
parameter no
longer displays.
Summary of Contents for Scalar 10K
Page 10: ...1 x Contents 6 00058 05 Rev A...
Page 22: ...2 xxii Figures 6 00058 05 Rev A...
Page 24: ...3 xxiv Tables 6 00058 05 Rev A...
Page 26: ...1 2 About This Guide 6 00058 05 Rev A...
Page 34: ...1 10 About This Guide 6 00058 05 Rev A...
Page 36: ...2 2 System Description 6 00058 05 Rev A...
Page 49: ...Internal Components 2 15 September 2007 Figure 2 7 I E Station Magazines...
Page 51: ...Connectivity 2 17 September 2007 Figure 2 8 Indirect Fibre Channel Attachment...
Page 58: ...3 2 System Description 6 00058 05 Rev A...
Page 252: ...4 174 Menus and Commands 6 00058 05 Rev A...
Page 254: ...5 2 Processing Media 6 00058 05 Rev A...
Page 276: ...6 2 Error Messages 6 00058 05 Rev A...
Page 318: ...6 44 Error Messages 6 00058 05 Rev A...
Page 324: ...in 6 Index 6 00058 05 Rev A...