Service Menu
4-139
September 2007
Fiducial Test Dialog
Path:
Main Menu
Service
Diags
Scanner
Fiducial
Test
Use the Fiducial Test dialog to test the scanner for proper
operation. See Figure 4-209.
Figure 4-209
Fiducial Test Dialog
Parameters
Description
Test
1
reads the Insert/Eject Station
1 fiducial and displays the
found edge coordinates.
2
reads the Insert/Eject
Station 2 and displays the
found edge coordinates.
Xcnt
Displays the x edge position
count in tenths of a mm
Ycnt
Displays the Y edge position
count in tenths of a mm
Continue
Y
to continue running the test.
N
to cancel the test
Run this test several times to ensure the X and Y position
counts are within 10 tenths of a mm each time the test is
run.
Loops:00 Test: 1
Xcnt:000000
Ycnt:000000
Continue: Y<
Summary of Contents for Scalar 10K
Page 10: ...1 x Contents 6 00058 05 Rev A...
Page 22: ...2 xxii Figures 6 00058 05 Rev A...
Page 24: ...3 xxiv Tables 6 00058 05 Rev A...
Page 26: ...1 2 About This Guide 6 00058 05 Rev A...
Page 34: ...1 10 About This Guide 6 00058 05 Rev A...
Page 36: ...2 2 System Description 6 00058 05 Rev A...
Page 49: ...Internal Components 2 15 September 2007 Figure 2 7 I E Station Magazines...
Page 51: ...Connectivity 2 17 September 2007 Figure 2 8 Indirect Fibre Channel Attachment...
Page 58: ...3 2 System Description 6 00058 05 Rev A...
Page 252: ...4 174 Menus and Commands 6 00058 05 Rev A...
Page 254: ...5 2 Processing Media 6 00058 05 Rev A...
Page 276: ...6 2 Error Messages 6 00058 05 Rev A...
Page 318: ...6 44 Error Messages 6 00058 05 Rev A...
Page 324: ...in 6 Index 6 00058 05 Rev A...