Service Menu
4-127
September 2007
Move Dialog
Path:
Main Menu
Service
Diags
Accessor
Move
Use the Move dialog to move the accessor in a figure eight
pattern without accessing cartridges. See Figure 4-190.
Figure 4-190
Move Dialog
Parameters
Description
Cycles to Run
Indicates the desired cycle
count (001–999)
Accept
Y
to accept changes
N
to reject changes
If
Y
is selected, the Response Screen appears.
See
Figure 4-191 on page 4-127. Otherwise, the changed
parameters continue to display but no action is taken.
Figure 4-191
Response Dialog
Parameters
Description
Cancel
Y
to cancel the test.
N
to continue running the test.
Cycle
Displays the number of
completed test cycles of the
requested cycles
Status
Displays the status of the
selected cycle (Running,
completed, Error, Cancelled)
Cancel
Y
to cancel the test.
N
to continue running the test.
If
Y
is selected for the Cancel parameter value, the test is
canceled. Otherwise, the test continues.
The accessor test
does corner moves.
Cycles to run:009<
Accept: N
DIAGS ACCESSOR TEST
Cycle :006 of 009
Status:Running
Cancel: N<
Note
If the operation
stops, the
Cancel
parameter no
longer displays.
Summary of Contents for Scalar 10K
Page 10: ...1 x Contents 6 00058 05 Rev A...
Page 22: ...2 xxii Figures 6 00058 05 Rev A...
Page 24: ...3 xxiv Tables 6 00058 05 Rev A...
Page 26: ...1 2 About This Guide 6 00058 05 Rev A...
Page 34: ...1 10 About This Guide 6 00058 05 Rev A...
Page 36: ...2 2 System Description 6 00058 05 Rev A...
Page 49: ...Internal Components 2 15 September 2007 Figure 2 7 I E Station Magazines...
Page 51: ...Connectivity 2 17 September 2007 Figure 2 8 Indirect Fibre Channel Attachment...
Page 58: ...3 2 System Description 6 00058 05 Rev A...
Page 252: ...4 174 Menus and Commands 6 00058 05 Rev A...
Page 254: ...5 2 Processing Media 6 00058 05 Rev A...
Page 276: ...6 2 Error Messages 6 00058 05 Rev A...
Page 318: ...6 44 Error Messages 6 00058 05 Rev A...
Page 324: ...in 6 Index 6 00058 05 Rev A...