Chapter 2 Assembly & Installation
Revision W
HF Series X-ray Generators - Service Manual
2-44
Quantum Medical Imaging
5. Position the OCP on the Wall Mount bracket aligning the mounting holes
in the bottom of OCP with the mounting holes in the wall bracket.
6. Secure the OCP to the Wall Mount Bracket using four (4) 1/4-20 x 3/4"
pan head phillips screws, flat washers and nuts.
7. Re-install four (4) OCP front panel screws.
8. Connect one end of the OCP serial cable (P/N AY51-018) to connector J3
on the OCP. Route the OCP serial cable through the access opening in
the Wall Mount Bracket and connect to connector J5 on CPU Board A1
(P/N AY40-006T). Ensure the connector screws are fully tightened. DO
NOT route serial cable through an opening containing a High Voltage
Cable (see Figure 2-13).
Q-VISION Operator Control Panel
The Q-VISION Operator Control Panel (OCP) is an all-in-one PC, with the
computer integrated with the touch-screen monitor in a single case. The Q-
VISION OCP is provided with a Power/Expose Control Box (AY20-271), which
contains a two-stage, hand-held PREP/EXPOSE switch and a built-in genera-
tor ON/STANDBY rocker switch for remotely powering on/off the Generator.
Cables for interconnection of the Q-VISION OCP and Power/Expose Control
Box to the generator are available in either 50-ft. or a 100-ft. interconnect
lengths, as specified at the time of order. In order to interface Q-VISION with
the generator, the generator must be equipped with the "Q-Connect" option,
which provides an OCP Interface Board (AY40-073S) that enables Ethernet
connectivity to the generator’s Logic Board AY40-006S.
The Q-VISION OCP monitor is intended for use as a "touch-screen" input
device; however, a keyboard and mouse are provided for use by field service
technicians if and when it becomes necessary to perform operations external
to the Q-VISION program environment (such as when performing operating
system functions).
The Q-VISION OCP can be placed directly on a flat surface such as a table top
or desk, or on can be wall-mounted using Wall Mount option QGV-WM. Refer
to the Q-Connect Option System Interconnection Diagram in Chapter 6, DIA-
GRAMS and proceed as follows:
1. Connect AC power adapter plug to power jack on the All-in-One PC (Q-VISION) in
accordance with manufacturer’s instructions (provided with PC).
2. If optional keyboard and/or mouse are provided, connect these in accor-
dance with manufacturer’s instructions (provided with PC).
3. Connect Ethernet cable from All-in-One PC (Q-VISION) to an Ethernet
Switch provided with Q-VISION OCP.
4. Connect Ethernet cable EL75-058-XXX between Ethernet Switch and Gen-
erator OCP Interface Board (AY40-073S) connector J1 (see Figure 2-24a).
5. Install the Ethernet Switch Power Supply connector, and plug transformer
into 110 VAC, 60Hz electric outlet (220 VAC, 50Hz power source applica-
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...