Appendix D Calibration (Using Generator Service Tools)
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging, LLC
D-53
•
KV - Calibrates the tube voltage settings used in conjunction with a
particular Receptor/Film Screen and ion chamber combination
•
Copy - The Copy Receptor Calibration utility enables stored AEC
calibration data (kV curves, AEC offset data) to be copied from a
calibrated receptor/film screen to another receptor/film screen in order
to reduce overall calibration time.
IMPORTANT! When calibrating AEC using a wall
stand mounted receptor, ensure that the x-ray
beam is perpendicular to image receptor and that
the correct ion chamber detector(s) is selected.
Proceed as follows:
1. Factory Reset check box - when selected, will reset AEC Calibration
values to factory default settings. This normally should not be selected;
it should only be selected when calibration values are extraordinarily out-
of-range (usually signifying a hardware or software abnormality).
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...