![Quantum ODYSSEY HF Series Service Manual Download Page 476](http://html1.mh-extra.com/html/quantum/odyssey-hf-series/odyssey-hf-series_service-manual_3538349476.webp)
Appendix D Calibration (Using Generator Service Tools)
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging, LLC
D-49
8. Repeat above steps with UPPER RIGHT ion chamber detector selected
and UPPER LEFT and UPPER CENTER detectors covered with lead. The
measured mAs reading should be within 5% of the mAs reading noted in
the previous step. If not, adjust the UPPER RIGHT detector gain potenti-
ometer, re-take exposure and verify it is within tolerance. Repeat as nec-
essary.
9. Access the Generator Settings Screen and reset Backup MAS to original
setting noted previously.
Note: If the system is equipped with
"universal" type AEC (i.e., AEC Board
A11 is part number AY40-027S), and
the ion chamber is "level" type (i.e.,
Siemens), adjust the appropriate detec-
tor gain potentiometer on AY40-027S in
accordance with the table below:
Table D-4. Universal AEC Board A11 Detector Gain
Adjustment Potentiometers (Siemens Ion Chamber Only)
ION CHAMBER
DETECTOR
AEC BOARD A11
(AY40-027S)
POTENTIOMETER
TABLE
1
R61
TABLE
2
R63
TABLE
3
R66
WALL#1
1
R54
WALL#1
2
R55
WALL#1
3
R56
WALL#2
1
R53
WALL#2
2
R58
WALL#2
3
R57
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...