Chapter 5 Service Instructions
Revision W
HF Series X-ray Generators - Service Manual
5-30
Quantum Medical Imaging, LLC
MA
Current
Indicates the exposure mA current
is out of range.
1. Verify that generator is properly
calibrated.
2. Check if current on standby is
approximately 2.5 amps. If not,
replace KVP Control Board A2
(AY40-003S). If problem persists,
replace Filament Control Board A5
(AY40-007T). Replace Logic
Board A1 (AY40-006S) if problem
is still not corrected.
3. On KVP Control Board A2, con-
nect oscilloscope probe to test
point A2TP7 to verify correct tube
current waveform. The current
waveform should indicate 1V/100
mA. If waveform is close to mA
setting, re-calibrate generator. If
waveform is zero, replace KVP
Control Board A2. If problem per-
sists, replace H.V. Transformer
A17 (AY20-031 on 125 kVp sys-
tem, or AY20-032 on 150 kVp sys-
tem).
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...