Appendix B Calibration (DiRex System)
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging, LLC
B-41
WALL STAND 40" SID
FILM/SCREEN
1. Copy "Wall 72" film/screen to "Wall 40".
2. Perform "AEC Optical Density (OD) Calibration",
however,
DO NOT
re-adjust the ion chamber
MASTER GAIN potentiometer. Adjust "AEC Ref"
to obtain desired (set) density.
3. Perform "AEC KV Calibration".
TABLE 40" SID FILM/
SCREEN
1. If system includes a wall stand, copy "Wall 40"
film/screen to "Table 40".
2. Perform "Ion Chamber Balancing" procedure for
table ion chamber.
3. Perform "AEC Optical Density (OD) Calibration":
a. Set desired "base" optical density
b. Set "AEC Ref" to 1200.
c. Adjust ion chamber’s MASTER GAIN potenti-
ometer until base density is achieved in
acquired images.
4. Perform "AEC KV Calibration".
Table B-4. CR System AEC Calibration Procedure Sequence
FILM/SCREENS TO BE
CALIBRATED
REQUIRED AEC CALIBRATION PROCEDURES/
SPECIAL INSTRUCTIONS
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...