Appendix D Calibration (Using Generator Service Tools)
Revision W
HF Series X-ray Generators - Service Manual
D-38
Quantum Medical Imaging, LLC
Advanced Filament Calibration Wizard
The "Advanced Filament Calibration" wizard is provided to improve sys-
tem accuracy at a selectable technique (i.e., kVp, mA and filament size).
Note that Advanced Filament Calibration is not required in order to
achieve satisfactory generator performance. It should only be performed
on techniques where mA accuracy or filament offset waveform optimiza-
tion is desired.
To access the Advanced Filament Calibration wizard, proceed as follows:
1. From the Calibration Wizard screen, select Tools>Advanced>Fila-
ment. The Calibration Wizard - Advanced Filament Calibration Wiz-
ard Intro 1 Tab is displayed:
Figure D-23. Calibration Wizard - Advanced Filament Calibration
Wizard Intro 1 Tab
2. Factory Reset check box - When selected, will reset Filament Offset
Calibration values to factory default settings. This normally should
not be selected; it should only be selected when calibration values
are extraordinarily out-of-range (usually signifying a hardware or
software abnormality).
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...