Appendix B Calibration (DiRex System)
Revision W
HF Series X-ray Generators - Service Manual
B-6
Quantum Medical Imaging, LLC
7. Upon validation of the password, the Service Options Menu displays:
Figure B-2. Service Options Menu
The Service Options Menu offers the following selections:
• Generator Service - Accesses the generator configuration, generator
calibration, Factory Reset, and Statistics utilities.
• Remoting Service - Accesses the Device Settings and System Settings
utilities. (These utilities are described in detail in the DiRex Service
Manual DC30-088.)
• Exit Application - Exits the Quantum generator OCP software
application (the Main OCP Overlay Menu will no longer appear on the
Workstation Computer)
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...