Quantum Medical Imaging
60950 for Information Technology Equipment, the workstation must be
placed outside the radius of the patient environment, shown in Figure 2-1.
Only devices conforming to UL 60601-1/IEC 60601-1 may be placed within
this environment. Consult with the user of the system when determining the
location for the workstation computer; consider all areas the patient may
access, including the room entrance and/or exit, when establishing the
patient environment.
Figure 2-1. Patient Environment
2.5 m
1.5 m
1.5 m
1.5 m