Battery
Over-
Charged
Battery 2
Over-
Charged
Line Monitor/Charger board not
adjusted correctly.
1. Connect a DMM between Line
Monitor/Charger Board AY40-
028T test point TP2 and ground.
Verify voltage is +204 to +206
VDC. If not, adjust A8R70 to
obtain correct voltage. If voltage
cannot be adjusted to correct
value, replace Line Monitor/
Charger Board A8.
2. Connect a DMM between test
point TP3 and ground. Verify volt-
age is -204 to -206 VDC. If not,
adjust A8R66 to obtain correct
voltage. If voltage cannot be
adjusted to correct value, replace
Line Monitor/Charger Board A8.
Battery
NotRdy
Battery voltage is too low. (This
warning message applies only to
stored energy generators.)
Wait until READY light on Operator
Control Panel is steady before taking
exposure.
Calibration
Generator kV, mA, and/or filament
calibration settings are outside of
acceptable limits for normal sys-
tem operation.
Access Service Mode and perform
kV, mA, and/or filament calibration.
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...