Appendix D Calibration (Using Generator Service Tools)
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging, LLC
D-39
3. Select Next to proceed to Advanced Filament Calibration Intro 2
Tab. Per the instructions provided on this screen, connect a Digital
Oscilloscope (Tektronix Model TDS2024 or equivalent) to test points
TP1 (channel 1), TP7 (channel 2), and TP4 (grounds) on KVP Control
Board A2 (AY40-003S) located on the generator PCB Door. Adjust the
Oscilloscope settings according to the screen instructions.
4. Select Next to proceed to Advanced Filament Calibration Intro 3
(Waveforms) Tab. The sample filament offset waveforms illustrate
three scenarios; filament offset adjusted too high, filament offset
adjusted too low, and correct filament offset adjustment. Keep these
in mind as you proceed through the Advanced Filament Calibration
process.
5. Select Next to proceed to Adjustment Point Selection Tab:
Figure D-24. Advanced Filament Calibration - Adjustment Point Selection Tab
6. If using an oscilloscope, select Scope radio button (Step 1.); if using
a mAs meter, select Mas Meter.
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...