Chapter 1 Specifications
Revision W
HF Series X-ray Generators - Service Manual
1-10
Quantum Medical Imaging
65 kW Three Phase (Model QG-65)
Line Voltage:
380-480 VAC (±10%), 90 Amps, 50/60 Hz
Momentary Current:
190 Amps (@440 VAC)
Long Term Current:
less than 1.0 Amp
Mode of Operation:
Intermittent operation
80 kW Three Phase (Model QG-80)
Line Voltage:
380-480 VAC (±10%), 100 Amps, 50/60 Hz
Momentary Current:
200 Amps (@440 VAC)
Long Term Current:
less than 1.0 Amp
Mode of Operation:
Intermittent operation
HIGH-SPEED STARTER POWER REQUIREMENTS
A High-Speed Starter (HSS) unit is optional on Models QG-40 and
QG-50, and is standard equipment on Models QG-65 and QG-80.
The external HSS (QG-HSS) configuration requires a separate 200 - 240 VAC, 20A
line input. Note that a separate 200- 240 VAC line input is
not required
on gen-
erators equipped with Internal High Speed Starter (Q-HSS) configuration.
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...