![Quantum ODYSSEY HF Series Service Manual Download Page 218](http://html1.mh-extra.com/html/quantum/odyssey-hf-series/odyssey-hf-series_service-manual_3538349218.webp)
Chapter 5 Service Instructions
Revision W
HF Series X-ray Generators - Service Manual
5-34
Quantum Medical Imaging, LLC
Not
Rotating
Systems equipped with a QMI Internal High-Speed Starter
:
1. Wrong system configuration.
1. Access the Service Menu>Config-
uration screen and set the Rotor
Type to "QMI HSS".
Check if jumper JP1 on HSS
Driver Board A20A2 (AY40-036S)
is installed. If not, install it.
2. Faulty HSS Control Board
A20A1 (AY40-035S).
2. Using oscilloscope, observe the
following signals on HSS Control
Board A20A1 (AY40-035S) (Note:
Synchronize the trigger on CH1
when it is falling.):
• CH1 - connect to U7-16 and
GND
• CH2 - connect to TP19 and GND
• CH3 - connect to TP14 and GND
Press PREP. Verify during PREP
the signal on TP19 is 0 Volts (low)
and TP14 is oscillating all the time.
If TP19 is oscillating, replace HSS
Control Board A20A1.
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...