Chapter 5 Service Instructions
Revision W
HF Series X-ray Generators - Service Manual
5-60
Quantum Medical Imaging, LLC
7. The Insight screen features seven expandable service utilities -
Receptor Info, Exposure Info, Lock Info, Direct Set, A/D Monitor,
Statistics, and Exposure log - for viewing and, where applicable,
changing settings of the Generator. Press the Statistics
expander. The Statistics Screen displays:
Figure 5-5f. Statistics Screen
The Statistics Screen provides access to various types of system
information. The following briefly describes the various Statistics
Screen selections:
•
Exposure Count: Indicates total number of exposures
taken on generator
•
Average KVP: provides mean KV value averaged over the
total exposures taken on generator
•
Average mA: provides mean mA value averaged over the
total exposures taken on generator
•
Average Time: provides the sum of exposure times (in
seconds) of all exposures taken on generator
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...