terminal blocks A11TB10, A11TB12 and A11TB11, respectively. Refer to Fig-
ure 2-21 for Siemens tomo interface cable wire connections. Secure the
tomo interface cable securely using cable ties.
Figure 2-21. Siemens Tomo/Ion Chamber Interface Cable Interconnection Diagram
Check that jumper JP1 is installed on Universal AEC Board A11 (AY40-027S).
If not, install a jumper.