![Quantum ODYSSEY HF Series Service Manual Download Page 222](http://html1.mh-extra.com/html/quantum/odyssey-hf-series/odyssey-hf-series_service-manual_3538349222.webp)
Chapter 5 Service Instructions
Revision W
HF Series X-ray Generators - Service Manual
5-38
Quantum Medical Imaging, LLC
Plus
15V
Missing or inc15 VDC sup-
ply voltage.
1. Access Service Mode>Calibra-
tion>AD Calibration menu. Verify
+15 VDC voltage is within toler-
ance per Table 3-8. If not, cali-
brate in accordance with "A/D
Calibration Procedure" in Chapter
3.
2. Check for +15 VDC (±10%) on
Power Supply Board A6 (AY40-
005T) test point TP10. If 0 VDC,
check fuse F3 on Interface Board
A9 (AY40-023T). If F3 is open,
replace fuse. If there still is no +15
VDC at A6TP10, replace Power
Supply Board.
3. If voltage at A6TP10 is not 0 VDC,
but is too low, check if another
board is shorting the +15 VDC out-
put. Disconnect output cable A6J2
and measure voltage at A6TP10. If
within range, check other boards
for shorts. If not within range,
replace Power Supply Board.
Plus
24V
Missing or inc24 VDC sup-
ply voltage.
1. Access Service Mode>Calibra-
tion>AD Calibration menu. Verify
+24 VDC voltage is within toler-
ance per Table 3-8. If not, cali-
brate in accordance with "A/D
Calibration Procedure" in
Chapter 3.
2. Check for +24 VDC (±10%) on
Power Supply Board A6 (AY40-
005T) test point TP8. If 0 VDC,
check fuse F4 on Interface Board
A9 (AY40-023T). If F4 is open,
replace fuse. If there still is no +24
VDC at A6TP8, replace Power
Supply Board.
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...