Appendix D Calibration (Using Generator Service Tools)
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging, LLC
D-23
•
Load From Generator - Select to load the last automatically-saved
generator configuration (Note: The configuration file is automatically
backed up every time the Workstation Computer establishes a
connection to the generator.) the currently active configuration (This
option should be selected under normal circumstances; only when a
problem occurs with the system should the Load From File option be
selected).
5. Select Load From Generator. The Calibration Wizard - Overview tab dis-
plays:
Figure D-13. Calibration Wizard - Overview Tab
The Calibration Wizard provides a sequence of dialog boxes that leads the
calibrator through a series of generator calibration steps, performed in a spe-
cific sequence. This wizard may be helpful especially to those who may be
unfamiliar with this equipment. The "Calibration Selection" pane, shown
in the sample screen above, identifies the calibration procedures to be per-
formed as indicated by a check mark next to each procedure. The check
mark is an active field; removing a check mark will eliminate that procedure
from the performance sequence, adding a check mark includes it.
CALIBRATION
SELECTION PANE
CHECK MARK =
PERFORM
NO CHECK MARK =
DO NOT PERFORM
CHECKED = DONE
UNCHECKED = NOT DONE
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...