Chapter 5 Service Instructions
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging, LLC
5-93
2. Place the new High-Speed Starter Driver Board A20A2 on the PCB
Door standoffs and secure using four (4) hex nuts.
3. Connect all cables to High-Speed Starter Driver Board A20A2.
4. Place clear plastic high-voltage shield on PCB Door standoffs and
secure using three (3) 6-32 hex nuts.
5. Set the ON/OFF circuit breaker CB1 on the generator cabinet to
OFF position.
6. Set the power on/standby switch on the Operator Control Panel
(OCP) to the standby ( ) position.
Stored Energy System - Battery Tray Removal
Figure 5-8. Battery Tray Component Location
Caution! These batteries contain Sulfuric Acid
(H
2
SO
4
) and Lead (Pb). Spent Lead-Acid batteries
must never be disposed of indiscriminately. QMI
requires its dealers to collect and send them to a
secondary lead smelter for recycling. In the U.S.,
any reseller of automotive-type storage batteries
is also required to accept spent cells at no cost to
you. Avoid damage to the battery casing. Keep
away from fire, flame, sparks, and heat. Use care
in handling.
Warning! Hazardous voltages (400 VDC) exist
inside the Battery Tray Assembly even when
power is disconnected from the generator. Use
caution when handling; do not put hands/fingers
inside Battery Tray Covers.
BATTERY TRAY
BRACKET (2X
ON 30/40 kW
SYSTEMS)
FUSE HOLDER
(2X ON 30/40
kW SYSTEMS)
CONNECTOR
A20TB1
CONNECTOR
A21TB1
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...