Chapter 2 Assembly & Installation
Revision W
HF Series X-ray Generators - Service Manual
2-36
Quantum Medical Imaging
Digital receptor cable entrance/exit is provided through the access opening in
the top rear panel of the generator chassis. DO NOT route signal cables
through an opening containing a High Voltage Cable (see Figure 2-13). When
connecting to a table digital receptor, remove jumper JP2 on Interface Board
A9. When connecting to wall digital receptor, remove jumper JP3 on Inter-
face Board A9.
Interfacing to CMT Digital Radiographic Receptor (When
Applicable)
Refer to CMT digital receptor installation manual for detailed wiring instruc-
tions.
Interfacing to Siemens Ion Chambers (When Applicable)
Note: When interfacing the generator to a Siemens table/wall
ion chamber and a Quantum table/wall ion chamber (AID type),
only the Siemens ion chambers require the use of cable
assembly AY51-061. The Quantum table/wall ion chamber can
be connected directly to AEC Board A11 (AY40-027S) connector
J10 (TABLE) or J8 (WALL).
Interfacing of the HF Series x-ray generator to Siemens table and/or wall ion
chambers, such as in Siemens Tomography System, is facilitated using the
"universal" type AEC Board A11 (AY40-027S) located on the front side of the
generator cabinet PCB door. The VERTIX wall stand and MULTIX table ion
chambers are connected as shown in Figure 2-19. Note that cable assemblies
AY51-061 shown in the illustration should already be connected to both the
MULTIX table AEC cable EK14 and the VERTIX wall stand AEC cable. Any
additional ion chambers used with the system are connected as described in
"Table/Wall Ion Chamber Connections (AEC Option Only)" procedure included
in this chapter.
Figure 2-19. AEC Ion Chamber Connections (Siemens Tomography System)
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...