![Quantum ODYSSEY HF Series Service Manual Download Page 456](http://html1.mh-extra.com/html/quantum/odyssey-hf-series/odyssey-hf-series_service-manual_3538349456.webp)
Appendix D Calibration (Using Generator Service Tools)
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging, LLC
D-29
3. Select Next to proceed with the first (50) kVp calibration, or you may
select a particular kVp Calibration to perform by selecting its correspond-
ing tab (choices are 50 kVp and 100 kVp). For this example, "50 kVp" is
selected, as shown below:
Figure D-17. Sample kVp Calibration Screen (50 kVp is shown)
4. Connect a kVp bleeder tank in series between X-ray tube and the genera-
tor’s H.V. Transformer (A17), or use a non-invasive kVp meter.
5. Take an exposure at the setting indicated on the 50kVp Calibration
Screen (i.e., 50 kVp, 200L mA, 100 msec).
6. On the kVp meter, observe the measured kVp value. Enter the actual
(measured) value by adjusting the slide control, or by selecting the data
entry box and typing the value using a keyboard. The value entered
should be to the hundredths decimal place. Press Calibrate button.
7. Ideally, the value appearing in the 50 kVp Calibration Screen (Step 1.) will
match the reading on the kVp meter within allowable tolerance. If so,
press the Confirm button. The calibration value is saved and the system
advances to the 100 kVp Calibration setting.
ENTER KVP
READING FROM
METER
EXPOSURE
SETTINGS
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...