Chapter 2 Assembly & Installation
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging
2-37
Secure the ion chamber cables to the PCB Door using cable ties as shown in
Figure 2-20.
Figure 2-20. Siemens Ion Chamber Cable Management
Interfacing to Siemens Tomography System (When Applicable)
The Siemens tomo interface cable (supplied with the Siemens system) has
attached to it at one end three plugs identified as "KK1B G1,G2 X10", "KK1B
G1,G2 X11" and "KK1B G1,G2 X12". Plugs "X10" (TABLE), "X11" (WALL) and
"X12" (TOMO SWEEP TIME) will need to be cut off and the wire ends stripped
to allow connection to the universal AEC Board A11 (AY40-027S) pluggable
SECURE ONE ION
CHAMBER CABLE
CONNECTOR HERE
SECURE OTHER
ION CHAMBER
CABLE CONNECTOR
HERE
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...