1.J. Environmental Requirements (Operating, Shipping and Storage Conditions)
Power Requirements:
100–240VAC, 50/60Hz, 2.5A
Temperature:
4°C–50°C (shipping/storage), 15°C–25°C (operation)
Humidity:
Up to 80% relative humidity, non-condensing
Operating Altitude:
<2000 meters
The Maxwell
®
RSC Instrument is intended for indoor use only. Wipe up spills immediately. To avoid
shortening the expected lifespan of the instrument, install in a location that meets the following criteria:
• Locate on a sturdy, level surface.
• Avoid dusty areas.
• Choose a location that has good air circulation and is not exposed to direct sunlight.
• Avoid electrically noisy power sources (e.g., power generators).
• Do not install in a location where there is large temperature variability or high humidity.
• Do not position the instrument so that it is difficult to unplug from the power source.
• Do not place next to heat sources.
• Do not use near flammable gases or liquids.
• Do not place near other electrically sensitive instruments.
• Connect the Maxwell
®
RSC Instrument and Tablet PC to a power outlet. Promega recommends that you
plug the instrument and tablet PC into an uninterruptible power supply.