Protocol Completes Successfully
When the protocol completes successfully the Protocol Selected window will change to indicate that the
protocol has been completed (Figure 45). After a protocol is complete the Current Step will be listed as
Completed
. Touch the
Open Door
button to open the door of the Maxwell
®
RSC Instrument.
Figure 45. Protocol Selected window after completion of the protocol.
Upon completion of the
currently running protocol, the Protocol Selected window changes to display a green title bar and 100%
completion in the run progress bar. The Current Step changes to Completed after a run is finished. After a
protocol is completed you can touch the
Open Door
button to open the door of the Maxwell
®
RSC
Instrument to remove the deck tray.
Close the caps of the elution tubes, and remove the tubes from the tray (Figure 46). Verify that all the
cartridges have a plunger in well #8. Remove the deck tray by gripping the tray firmly by the elution tube
position, lifting up the tray and pulling it out (see Figure 47). The deck tray may be warm to the touch after
a run is completed. Exercise caution while removing the deck tray. If there were cartridges missing plungers,
proceed to Section 5.C to remove them by running the Clean Up method. The extracted material is present
in the elution tubes. Remove the cartridges and plungers from the Maxwell
®
RSC Deck Tray.
Important:
Used cartridges and plungers should be disposed of appropriately according to your institution’s
procedures for hazardous and biohazardous waste. Do not reuse reagent cartridges, plungers or elution tubes.
12227T