5.B. Starting a Protocol (continued)
2. You will be presented with a screen that will ask you to select the method from the list matching the
kit name and or product number with that printed on the label on the kit box. Alternatively, you can
use an optional bar code scanner to scan the bar code on the Maxwell
®
RSC reagent kit box indicated
by the Scan Here label (Figure 35) or touch the text box (Figure 36) to manually enter the bar code
information using the keyboard. Maxwell
®
RSC requires manual entry of bar codes in the following
format: Product Catalog Number, Kit Lot Number, Expiration Date in year-month format (example:
AS13303221872015-05, where the product catalog number is AS1330, the kit lot number is 322187
and the expiration date is 2015-05). Only if the bar code entered has this format will the corresponding
processing method be selected and the
Proceed
button on the right side of the method become active.
Once the bar code is correctly entered, touch the
Proceed
button to get to the Cartridge Setup screen.
You can touch the
Back
button to return to the Home screen. Scanning or entering the bar code
information from a kit will select the method to be run (assuming it is installed in the software) and will
also record kit lot and expiration date information for the extraction run report.
Figure 35. Kit label indicating bar code to scan.
Shown in the blue box is the bar code to scan on the
kit label for starting a purification run.
Figur
e Number
12206T
A