Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
59
www.promega.com
TM411 · Revised 4/15
Figure 64. Sanitize Checklist screen.
The sanitization checklist informs you on how long the UV
sanitization will take. It will also ask you to confirm that no samples or eluates are present in the instrument
prior to running Sanitization.
After Sanitization is completed, you will see a flashing indicator:
Touch the
indicator
to view the Sanitization report.
Note:
UV radiation is useful in decontamination due to its ability to inactivate biological molecules.
UV treatment is not a substitute for cleaning. Using the UV Sanitization protocol alone may not provide
sufficient decontamination.
Figure 65. Example Sanitization Report.
The sanitization report will be displayed after completion of
sanitization.
12234T
A
12235T
A
12216T
A