Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
39
www.promega.com
TM411 · Revised 4/15
• Multiple kit lots can be entered for a run by swiping across multiple black box positions and
scanning or entering additional kit bar code information (next to the Scan Here label). Touching
the arrow on the right or left side of the information entry area will move to the next available
cartridge position. When entering multiple kit lot numbers manually, the bar code needs to
conform to the format: Product Catalog Number, Kit Lot Number, Expiration Date in year-
month format (example: AS13303221872015-05 where the product catalog number is AS1330,
the kit lot number is 322187 and the expiration date is 2015-05). Note that positions with
incomplete information will display a red circle with an exclamation point at the top of the blue
rectangle. The
Proceed
button will be gray and inactive while required information is missing.
Once all information is correctly entered, touch the
Proceed
button to prepare the instrument for this
extraction run.
Important:
There are two bar codes on the kit label. When entering optional product and lot information,
use the data contained in the bar code at the top of the label. You will receive an error message if you enter
data from the wrong bar code on the kit.
Figure 38. Cartridge Setup screen.
At the top of this screen the selected protocol is indicated. This
screen allows the user to select which cartridge positions will be processed. To select/deselect a
cartridge position touch the blue rectangle for any given position.
!
12223T
A