LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
NIBP Connector Replacement
308
8
Installing the NIBP Connector
Refer to
To install the NIBP connector into the parameter bezel (rear case—refer to
Inside Rear Case Diagrams (p. 228)
NOTE:
When installing a new NIBP connector, use
NIBP Connector Repair Kit (MIN 3305431-022) (p. 510)
.
1.
Install the connector (
) by installing the fitting nut on the back side of the bezel. Ensure the O ring seal is on the
NIBP connector (torque to 10.0 in-lb using a 1/2" deep socket) (see
for NIBP connector orientation).
2.
).
NOTE:
If the NIBP tube has been previously disconnected/reconnected from the fitting, the tube end should be trimmed
to maintain an airtight seal.
3.
Install the parameter bezel as described in
Installing the Parameter Bezel (p. 313)
.
4.
Install the system/therapy PCB assembly as described in
Installing the System (A01)/Therapy (A04) PCB Assembly (p. 239)
5.
Reassemble the case as described in
Reassembling the Case (p. 184)
.