LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
NIBP (A21)/CO2 (A23) Module Replacement
290
8
11. For the NIBP option disassembly, remove the NIBP cable (W27) from cable
guide on bracket cover.
) and lift off the cover. Discard the screws.
13. To remove the NIBP sub-assembly (
2
,
remove the four screws (
) and lift the assembly off the OEM
bracket. Discard the screws.
14. Disconnect the OEM PCB/NIBP module cable (W27) from the A21 NIBP
PCB at J2.
To remove the NIBP and CO2 modules as a unit from the rear case:
(Continued) 17 steps, (Page 3 of 4)
cover screws
(4 places)
Figure 8.83—Bracket cover screw locations
NIBP option only
screws (4 places)
Figure 8.84—NIBP sub-assembly screw locations