LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
SpO2 PCB (A16) Replacement
283
8
8.
Route CO2 adapter cable (W30) above the power/system cable
as shown.
9.
Install power/system cable (W01) to the power board at J17.
10. Install the OEM shield (
connector and the J11 and J15 connectors.
11. Install the power/therapy cable (W02) (
) to J8 on
the power PCB.
12. Install the system/therapy PCB assembly as described in
System (A01)/Therapy (A04) PCB Assembly Replacement (p.
232)
.
13. Reassemble the case as described in
.
To install the SpO2 Module in the rear case (see
Figure 9.21 on p. 413
):
(Continued) 13 steps, (Page 4 of 4)
Figure 8.76—Power/system PCB connections
J17
connection
J8
connection
CO2
adapter
cable
standoff
locations
Figure 8.77—OEM shield
OEM shield