LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
OEM PCB (A06) Replacement
266
8
4.
Install the nylon standoff (
the OEM hole (torque to 4.0 in-lb using a
1/4” socket).
5.
If present, connect the SpO2/OEM cable
(W21) (
feature is engaged.
6.
If present, connect the CO2/OEM cable
(W26) (
feature is engaged.
7.
If present, connect the NIBP/OEM cable
(W27) (
).
To install the OEM PCB (A06) into the rear case:
(Continued) 12 steps, (Page 2 of 3)
Nylon
standoff
W27 - NIBP /
OEM cable
W26 - CO2/
OEM cable
W21 - SpO2/
OEM cable
Figure 8.63—OEM PCB connections