LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Installing the System (A01)/Therapy (A04) PCB Assembly
242
8
10. Line up the system PCB with the inserts in the rear case, and
secure with seven new screws (
); torque to 6.8 in-lb.
11. Capture the ground strap from the EMI Shield with one new screw
); torque to 6.8 in-lb.
To install the system/therapy PCB:
(Continued) 15 steps, (Page 4 of 5)
system/therapy mounting screws
(7 places)
mounting screw with washer
(1 place)
Figure 8.43—System PCB screws