LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Biphasic Module (A22)/Inductive Resistor (A14) Replacement
300
8
6.
Remove the two biphasic module mounting screws (
).
Discard the screws.
7.
Lift the biphasic bracket up slightly to gain access to connector. Disconnect
the red transfer relay lead spade connector from the biphasic modular
assembly at J104. Needle nose pliers may be used to assist in
disconnection.
8.
Remove the bracket (
) with the biphasic PCB (
) and inductive resistor (
9.
) can be removed from the bracket prior
to removing the biphasic PCB from the bracket. To remove the inductive
resistor from the bracket disconnect the two spade terminal connectors at
J102 and J108 of the biphasic PCB. Remove two screws (
)
from the biphasic bracket, lift the mounting tabs, and lift the inductive
resistor from the assembly. Discard the screws.
10. To continue removing the biphasic PCB, remove the screw (
)
and ground strap
) from the PCB. Discard the screw.
Figure 8.95—Biphasic bracket and inductive resistor
To remove the biphasic module and/or the inductive resistor from the rear case (refer to
Inside Rear Case Diagrams (p. 228)
)
:
(Continued) 12 steps, (Page 2 of 3)
biphasic bracket,
mounting screws
with washers
(2 places)
ground strap, screw
with washer
J102
J108
inductive
resistor
screws and
washers
(2 places)