LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Installing the System (A01)/Therapy (A04) PCB Assembly
247
8
Separating the Therapy PCB (A04)
To remove the therapy PCB (A04):
1.
Disassemble the case as
described in
2.
Remove the system/therapy
assembly as described in
Removing the System (A01)/
Therapy (A04) PCB Assembly (p.
233)
.
3.
Disconnect the BTE cable (W20)
at the P51 connector from the
therapy PCB assembly.
Figure 8.47—P51 therapy PCB connection
W20 BTE
cable J51