LIFEPAK 15 Monitor/Defibrillator
Service Manual
Preventive Maintenance
Scheduled Replacement Items
149
6
Scheduled Replacement Items
The following items should be replaced at regular intervals because of their finite life span:
•
Disposable Electrodes
—Replace electrodes by the
“Use by” date.
•
ECG cable (3-lead/5-wire/12-lead)
—Replace to ensure the continued performance of this cable.
•
Lithium-ion battery pak
— Replace to ensure maximum operating time of the device.
•
Coin battery
— Replace to ensure the device will not lose battery power for the real-time clock.
NOTE:
Contact your Physio-Control Service Representative for assistance when coin battery replacement is required.
shows the schedule for replacement items.
Table 6.2—Replacement Schedule
Replacement Item
Frequency
Replace ECG cable
2 years
Replace battery pak
2 years
Replace coin (clock) battery
5 years