LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
NIBP (A21)/CO2 (A23) Module Replacement
295
8
4.
For the NIBP option, route the NIBP cable through the cable mount on
side of bracket cover.
5.
Place the OEM bracket into rear case and secure with the three new
mounting screws (
To install the A21 NIBP, A23 CO2, and SpO2 PCBs:
(Continued) 12 steps, (Page 4 of 6)
NIBP cable mount
Figure 8.90—NIBP cable routing
bracket mounting
screws (3 places)
CO2
exhaust
from rear
case
Figure 8.91—CO2/NIBP mounting bracket