LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Front Case Replacement
210
8
Assembling the Front Case
Refer to
To install the new front case:
NOTE:
Front Case Repair Kit (MIN 3305431-023) (p. 511)
NOTE:
Transfer the following parts from the old front case to the new front case:
~ LCD/interface PCB assembly
~ W15 speed dial assembly
~ W17 speaker assembly
~ W11 therapy connector
~ W16 printer/interface PCB cable
~ W12 printer control keypad cable
~ W13 main keypad cable
~ Printer bracket (
1.
) into the groove on the front case.
NOTE:
Thread the perimeter seal approximately 1.5 inches through the seal lock.
Do not stretch the seal, as this will result in crimped corners.
2.
Install the speed dial assembly as described in
Installing the Speed Dial Assembly (W15)
.
3.
Install the therapy connector cable as described in
Installing the Therapy Connector Cable
4.
Install the printer assembly/interface PCB Cable (includes the W19 cable) as described in
Installing the Printer Assembly/Interface PCB Cable (W16) (p. 222)
.