LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Therapy Connector Cable (W11) Replacement
214
8
Therapy Connector Cable (W11) Replacement
Refer to
Inside Front Case Diagram (p. 189)
Removing the Therapy Connector Cable (W11)
To remove the therapy connector cable (see
) from the front case:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
From the outside of the front case, remove the four screws (
the therapy connector.
3.
Lift the therapy Connector cable (
) and therapy mounting seal from
the front case. Discard the four screws and therapy mounting seal (
).
Installing the Therapy Connector Cable (W11)
To install the therapy Connector cable into the front case:
NOTE:
If therapy connector cable requires replacement, use the
Therapy Connector Repair Kit (MIN 3305431-009) (p. 503)
1.
From the outside of the front case, install a new therapy mounting seal
(
)
into the front case.
CAUTION
POSSIBLE MOISTURE LEAKAGE
When installing the therapy connector, use a new
) to help prevent ingress of fluids.
Figure 8.23—Therapy connector cable
and seal