LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
CO2 Adapter Cable (W30) Replacement
354
8
Removing the CO2 Adapter Cable (W30)
Refer to
To remove the CO2 adapter cable (
) from the rear case:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as described in
Removing the System (A01)/Therapy (A04) PCB Assembly (p. 233)
3.
Remove the OEM PCB (A06) as described in
Removing the OEM PCB (A06) (p. 262)
.
4.
Remove the NIBP/CO2 module as described in
Removing the NIBP (A21)/CO2 (A23) Modules (p. 288)
.
5.
Disconnect the CO2 adapter cable from the CO2 PCB (A23) at J1.
Installing the CO2 Adapter Cable (W30)
Refer to
To install the CO2 adapter cable (
) from the rear case:
1.
Connect the CO2 adapter cable to the A23 CO2 PCB at J1.
2.
Install the NIBP/CO2 Module as described in
Installing the NIBP (A21)/CO2 (A23) Modules (p. 292)
.
3.
Install the OEM PCB (A06) as described in
Installing the OEM PCB (A06) (p. 265)
.
4.
Install the system/therapy PCB assembly as described in
Installing the System (A01)/Therapy (A04) PCB Assembly (p. 239)
5.
Reassemble the case as described in
Reassembling the Case (p. 184)
.