LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Reassembling the Case
184
8
Reassembling the Case
To reassemble the case halves:
(Page 1 of 5)
1.
Position the front case over the rear case.
2.
Connect the P23 therapy ribbon cable connector (W11)
(
) from the front case to J23 of therapy PCB
).
3.
Connect the J24 6-pin block therapy cable connector (W11) to P24
transfer relay connector (A13).
Connect P23
ribbon cable
Connect J24
therapy cable