LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
NIBP (A21)/CO2 (A23) Module Replacement
288
8
Removing the NIBP (A21)/CO2 (A23) Modules
To remove the NIBP and CO2 modules as a unit from the rear case:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as described in
Therapy (A04) PCB Assembly Replacement (p. 232)
.
3.
Remove the transfer relay assembly as described in
.
NOTE:
It is not necessary to remove the spade terminals as
described in the procedure.
4.
Remove the OEM PCB as described in
Removing the OEM PCB (A06) (p.
.
5.
Remove the three mounting bracket screws (
screws.
6.
If present, disconnect the CO2 exhaust tube. Lift the assembly as needed.
bracket mounting
screws (3 places)
CO2 exhaust from
rear case
Figure 8.81—CO2/NIBP mounting bracket