LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
NIBP (A21)/CO2 (A23) Module Replacement
291
8
15. To remove the CO2 sub-assembly (
), remove the four screws
) and lift the assembly off the OEM bracket. Discard the
screws.
16. Disconnect the OEM PCB/CO2 module cable from the CO2 PCB (A23) at
J4.
17. Disconnect the CO2 adapter cable from the CO2 PCB (A23) at J1.
CAUTION
POSSIBLE SKIN BURNS
Do not open the CO2 scrubber device (part of the CO2 module). Scrubber material may cause
caustic burns. If scrubber material comes in contact with skin, rinse the area of contact thoroughly with water. If scrubber
material comes in contact with eyes, flush eyes with water for 15 minutes, and then seek immediate medical attention.
To remove the NIBP and CO2 modules as a unit from the rear case:
(Continued) 17 steps, (Page 4 of 4)
CO2 option only
screws (4 places)
Figure 8.85—CO2 sub-assembly screw locations