LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
System (A01)/Therapy (A04) PCB Assembly Replacement
233
8
Removing the System (A01)/Therapy (A04) PCB Assembly
To remove the system/therapy PCBs as a single unit from the rear case:
1.
Disassemble the case as described in
.
2.
Disconnect the connectors on the system PCB as follows:
~ J1 – Press the connector locking tabs and disconnect the
power/system PCB cable (W01,
).
~ J2 – (The system/interface PCB cable was disconnected
Disassembling the Case (p. 181)
~ J4 – Disconnect the USB flex connector (W14,
~ J6 – Disconnect the ECG connector cable (W7,
~ (If present) J7 – Disconnect the IP connector of the IP wire
).
~ (If present) J7 – Disconnect the temperature connector of
).
J1 power/
system PCB
cable
J4 USB flex
connector
J6 ECG cable
J7 IP/Temp cable
(if present)
Figure 8.35—System
PCB connectors