LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
System (A01)/Therapy (A04) PCB Assembly Replacement
238
8
8.
Lift the system/therapy board assembly out from the rear case.
Turn the system/therapy board assembly over to expose the
therapy board.
9.
Immediately discharge the pacing capacitor as described in
Discharging the C15 Pacing Capacitor (p. 179)
. The discharge
points are located at resistors R5 and R20 on the therapy board.
To remove the system/therapy PCBs as a single unit from the rear case:
(Continued) 9 steps, (Page 6 of 6)